Sensor Unit Unit
Step Select Input Adjust Verify Reading*
1 RTD-100 390.26P1** 1562.0°F±0.5°F
2 RTD-100 100.00Ω√32.0±0.1°F
3 RTD-100 18.49Ω√–328.0±0.2°F
4 RTD-1000 3,902.6P1** 1562.0°F±0.5°F
5 RTD-1000 1000.0Ω√32.0+0.1°F
6 1,000900.00P1 900.00
7 1,0001.00Ω√1.00±0.10
8 10,0009,000.0P1 9000.0
9 10,00010.0Ω√10.0±1.0
Table 3: Calibration of Meter-Mode Resistance Functions
(RTD-100/1000)
* exclusive of noise.
** adjust as necessary to obtain steps 2, 5.
Calibration Adjustments:
NOTE: Do not deviate from the calibration adjustment sequence
that follows. This will ensure that adjustments to be stored in
EEPROM go to the correct locations.
1. Perform meter-mode adjustments and verifications per Table 3 (use Table
5 for RTD-Thermistor models).
2. Hit OPR key (Flashing 'RTD' annunciator turns off).
3. Install J1 jumper.
4. Set unit to CALIB mode, RTD-100, °F.
5. Switch current source to OPERATE mode.
6 Perform calibrator-mode adjustments and verifications per Table 4 (use
Table 6 for RTD-Thermistor models).
7. Turn off unit. Remove tape holding battery.
8. Re-install back cover. Resistance-function calibration is complete.
30