Omega DP3001 General Information, Specifications Analog Input, Reference Input for 3-WIRE Ratio

Models: DP3001

1 24
Download 24 pages 8.6 Kb
Page 7
Image 7

1.0GENERAL INFORMATION

This voltmeter is a 4 1/2 digit panel meter for applications which require a compact, quality DC voltmeter. This model uses dual-slope conversion. Additionally, it provides automatic zeroing before each reading and does so with a minimum of parts for increased reliability.

The voltmeter has a resolution of 1 part in ±19999 counts. By using an external DC voltage reference, any of the given ranges can be configured for 3-wire (common ground) ratio measurement with readout from 0 to 1.9999.

Standard features include selectable decimal point location, display hold, test and blanking. Each model contains the required circuitry for full range selection and easy configuration.

2.0SPECIFICATIONS

2.1ANALOG INPUT

Model

-3

-4

-5

Range

±1.9999 Vdc

±19.999 Vdc

±199.99 Vdc

Resolution

0.1 mV

1 mV

10 mV

Maximum input

250 V rms, 350 Vp

250 V rms, 1000 Vp

250 V rms,1000 Vp

Input resistance

1 Gohm

1 Mohm

1 Mohm

Bias current

10 pA

1 pA

1 pA

Configuration

differential

 

 

Zero

automatic

 

 

Span adjustment

±5%

 

 

2.2REFERENCE INPUT FOR 3-WIRE RATIO

 

Analog input range

±2 Vdc, ±20 Vdc, ±200 Vdc

 

Reference voltage

+0.5 to +2.0 V

 

Load on reference

80 ohm (std), 100 Mohm (opt)

 

Accuracy

99.95%

2.3

NOISE REJECTION

 

 

NMR, sig hi to sig lo

56 dB at 50/60 Hz

 

CMR, sig lo to ana gnd

86 dB from DC to 60 Hz

 

CMV, sig lo to ana gnd

±1.0 Vdc

 

CMR, ac gnd to ana gnd

120 dB from DC to 60 Hz

 

CMV, ac gnd to ana gnd

2100 Vp per HV test, 354 Vp per IEC 348 spacing

1

Page 7
Image 7
Omega DP3001 manual General Information, Specifications Analog Input, Reference Input for 3-WIRE Ratio, Noise Rejection