1.0GENERAL INFORMATION
This voltmeter is a 4 1/2 digit panel meter for applications which require a compact, quality DC voltmeter. This model uses
The voltmeter has a resolution of 1 part in ±19999 counts. By using an external DC voltage reference, any of the given ranges can be configured for
Standard features include selectable decimal point location, display hold, test and blanking. Each model contains the required circuitry for full range selection and easy configuration.
2.0SPECIFICATIONS
2.1ANALOG INPUT
Model | |||
Range | ±1.9999 Vdc | ±19.999 Vdc | ±199.99 Vdc |
Resolution | 0.1 mV | 1 mV | 10 mV |
Maximum input | 250 V rms, 350 Vp | 250 V rms, 1000 Vp | 250 V rms,1000 Vp |
Input resistance | 1 Gohm | 1 Mohm | 1 Mohm |
Bias current | 10 pA | 1 pA | 1 pA |
Configuration | differential |
|
|
Zero | automatic |
|
|
Span adjustment | ±5% |
|
|
2.2REFERENCE INPUT FOR 3-WIRE RATIO
| Analog input range | ±2 Vdc, ±20 Vdc, ±200 Vdc |
| Reference voltage | +0.5 to +2.0 V |
| Load on reference | 80 ohm (std), 100 Mohm (opt) |
| Accuracy | 99.95% |
2.3 | NOISE REJECTION |
|
| NMR, sig hi to sig lo | 56 dB at 50/60 Hz |
| CMR, sig lo to ana gnd | 86 dB from DC to 60 Hz |
| CMV, sig lo to ana gnd | ±1.0 Vdc |
| CMR, ac gnd to ana gnd | 120 dB from DC to 60 Hz |
| CMV, ac gnd to ana gnd | 2100 Vp per HV test, 354 Vp per IEC 348 spacing |
1