PRODUCT INSTRUCTION SHEET
SECTION 8.0
CALIBRATION
IMPORTANT NOTE: SENSORS ARE SUPPLIED FACTORY CALIBRATED WITH A
PERIODIC CALIBRATION ( ABOUT ONCE PER WEEK) IS RECOMMENDED. THIS IS USEFUL IN TRACKING SENSOR FAILURES AS WELL.
8.1Removal of cord grip to access ZERO and SPAN potentiometers inside the sensor body:
a.Remove top nut of cord grip as shown in FIGURE 4.
b.Remove remaining cord grip as shown in FIGURE 5. A wrench may be required.
8.2Slope Calibration:
a.Determine the free chlorine content using a
b.Measure Free Chlorine content with sensor. Make sure that calibration flow rate and pH matches flow rate when measuring sample since probe output is flow rate and pH dependent. Make sure pH is within
c.Adjust the span potentiometer to the desired mA (see
FIGURES 9 & 9A)
c. Repeat this slope calibration one day after sensor is initially installed.
d. Repeat the slope calibration monthly.
SECTION 9.0
SENSOR STORAGE
9.0STORAGE. Store sensor at 5o C- 50o C ONLY and maximum humidity of 90%
a.Short Term Storage (1 week or less): Store in Flow cell with water to prevent the probe from drying out.
b.Intermediate Term (1 week to 1 month): Store in cap, bottle, or beaker with water to keep membrane wet.
c.Long Term (1 month or longer): Remove Membrane Cap and store completely immersed in tap water. Turn sensor upright and shake it to remove fill solution from inside the sensor.
d.Electrolyte bottle
FIG. 4 |
FIG. 6 |
FIG. 9 |
FreeC Cl2
LO
FIG. 5 |
FIG. 7 |
Left |
|
|
| Right | |||
potentiometer |
|
|
| potentiometer | |||
(Zero adjust) |
|
|
| (Span adjust) | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Cable
FIG. 8 |
to Increase output
to Decrease output
| ZERO READ | Note: The DPD method | ||
| for Free chlorine mea- | |||
|
|
|
| |
| FreeClO2Cl |
| ||
|
| surement is not selective | ||
|
|
|
| |
|
|
|
| for free chlorine alone. |
|
|
|
| |
|
|
|
| Other oxidants in the |
|
|
|
| water can also be mea- |
|
|
|
| sured as free chlorine. |
|
|
|
| M4679/0708 page 4 of 6 |
Parts covered by this product instruction sheet include: