Omega manual OS35-RS, Installation Instructions

Models: OS35

1 10
Download 10 pages 58.68 Kb
Page 7
Image 7

OS35

800-826-6342 (USA/Canada)

and OS35-RS

www.omega.com

 

(203) 359-1660 (International)

Infrared Temperature Sensor

INSTALLATION INSTRUCTIONS

The following procedure is recommended:

1.Install the SmartIRt/c as close as practical to view the target material to be measured.

2.Wire the SmartIRt/c to the controller, PLC, etc. in standard fashion (see PIN- OUT and wiring diagram)

3.Bring the process up to normal operating temperature and measure the actual temperature of the target material with the Microscanner D-Series, DX-Series Infrared Thermometer or any other reliable reference.

4.Adjust "input offset," "zero," "low cal," on the readout device to match the

reading of your reference.

Installation Complete. (For OEM installations preset the same adjustments. Indi- vidual calibration is not required.)

1

3

5

4

2

WIRING DIAGRAM

 

 

PIN OUT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Carol C074 Series/Belden 953 Series/Alpha 630 Series

 

 

 

Pin

Color

 

Function

 

 

Pin

Color

 

Function

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

White

 

 

Positive Output Signal

 

6

Brown

 

TA Pull Down Pin / IO option

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Green

 

 

Reference Output Signal

 

7

Yellow

 

RS232_TXD / IO option

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

Red

 

 

Positive Power

 

8

Violet

 

RS232_RXD / IO option

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

Black

 

 

 

Ground

 

9

Blue

 

RS232_RTD (READY_TO_SEND)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5

Bare

 

 

 

Shield

 

10

Orange

 

RS232_CTS (CLEAR_TO_SEND)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ORDERING INFORMATION

 

 

 

 

 

 

SmartIRt/c-F-P-O-T or Smart-microIRt/c-F-P-O-T (F is -. .SV, .4, .4SV only for Smart-microIRt/c)

 

 

F

FIELD OF VIEW

P

POWER

O

OUTPUT

T

TEMPERATURE RANGE

 

 

.3

3:1

 

12V

12 VDC

05

0-5V

100C

0-100C

 

 

.5

5:1

 

24V

24 VDC

010

0-10V

250C

0-250C

 

.20

20:1

 

 

 

 

420

4-20 mA

500C

0-500C

 

.40

40:1

 

 

 

 

232

RS232

1000C

0-1000C

 

-

1:2

 

 

 

 

 

 

 

70C

-30-70C

 

 

.SV

1:2

 

 

 

 

 

 

 

 

 

 

 

 

.4

4:1

 

 

 

 

 

 

 

 

 

 

 

 

 

.4SV

4:1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OPTION: Add Ta to any part # to add ambient temperature option to output ambient temperature surrounding sensor

NOTE: 0-10V only available in 24VDC and RS232 only available in 12VDC from 0-250C