HH 504 / HH 505 CALIBRATION PROCEDURE
Note: The following calibration procedure should perform only by qualified technicians
who have access to the items as following.
Equipment: The class of calibrator had better 10 times greater than the measured meter.
1. Turn off the meter then set the adjusted jumper to J1.
2. Insert the 383.00Ω calibrator to the T1 then turn on the meter, the main
display reading reads 1830.0 then press " ENTER " key.
3. The second display is stabilized (about 5sec.), then press " ENTER " key
when the meter auto power off then set the jumper back to J3 position.
4. Turn on the meter then Insert the 100.00Ω calibrator. The display reading
reads 0°C if the calibration procedure is right.
Pt385 Pt3926
-20018.521Ω16.996Ω
-10060.256Ω59.479Ω
0100.000Ω100.000Ω
100138.505Ω139.272Ω
200175.856Ω177.362Ω
300212.052Ω214.275Ω
400247.092Ω250.018Ω
500280.977Ω284.591Ω
600313.708Ω317.994Ω
700345.280Ω
800375.700Ω
383.00W CALIBRATOR
+RW1
RW1=RW2=RW3
RW=WIRERESISTANCE
G
+
_
_RW2 383.00W
GRW3