5. Handling Errors and Problems

5.1Error Messages

Error Display

Cause

Remedy

 

 

 

 

 

Press air release button and

 

Cuff is under inflated.

restart the measurement with a

 

higher inflation level. Carefully

 

 

 

 

read steps under section 3.3.

 

 

 

 

Movement during

Repeat measurement. Remain

 

still and do not talk during

 

measurement

measurement.

 

 

Refer to section 3.3.

 

 

 

 

Air tube disconnected.

Insert the air tube securely.

 

Refer to section 3.2.

 

 

 

 

 

 

Arm cuff not applied

Apply the arm cuff correctly.

 

correctly.

Refer to section 3.2.

 

 

 

 

Clothing is interfering

Remove any clothing

 

interfering with the arm cuff.

 

with the arm cuff.

 

Refer to section 3.2.

 

 

 

 

 

 

Air is leaking from the

Replace cuff with new one.

 

arm cuff.

Refer to Chapter 7.

 

 

 

 

The arm cuff was

Release your hand from the air

 

inflation bulb before the

 

inflated above

 

pressure reaches 299 mmHg.

 

299 mmHg.

 

Refer to section 3.3.

 

 

 

 

 

 

 

Replace all four “AAA”

 

Battery power is low.

batteries with new ones.

Blinks or

 

Refer to section 2.1.

 

 

appears

 

 

continuously

 

 

 

 

 