Omron M2 basic Handling Errors and Problems Error Messages, Error Display Cause Remedy

Models: M2 basic

1 34
Download 34 pages 34.38 Kb
Page 23
Image 23

5.Handling Errors and Problems

5. Handling Errors and Problems

5.1Error Messages

Error Display

Cause

Remedy

 

 

 

 

Cuff is under inflated.

Carefully read and repeat the

 

steps listed under section 3.4.

 

 

 

 

 

 

 

Repeat measurement. Remain

 

Movement during

still and do not talk during

 

measurement.

measurement.

 

 

Refer to section 3.3.

 

 

 

 

Air plug disconnected.

Insert the air plug securely.

 

Refer to section 3.2.

 

 

 

 

 

 

Arm cuff not applied

Apply the arm cuff correctly.

 

correctly.

Refer to section 3.2.

 

 

 

 

Clothing is interfering

Remove any clothing

 

interfering with the arm cuff.

 

with the arm cuff.

 

Refer to section 3.2.

 

 

 

 

 

 

Air is leaking from the

Replace cuff with new one.

 

arm cuff.

Refer to Chapter 7.

 

 

 

 

The arm cuff was

Do not inflate the arm cuff

 

inflated above 299

 

above 299 mmHg.

 

mmHg when inflating

 

Refer to section 3.4.

 

the cuff manually.

 

 

 

 

 

EN

23

Page 23
Image 23
Omron M2 basic instruction manual Handling Errors and Problems Error Messages, Error Display Cause Remedy