26
8.Technical Data
8. Technical Data

Note: Subject to technical modification without prior notice.

This device fulfils the previsions of EC directive 93/42/EEC (Medical Device Directive).

Product Description Digital Automatic Blood Pressure Monitor
Model OMRON M6 Comfort (HEM-7000-E)
Display LCD Digital Display
Measurement Method Oscillometric method
Measurement Range Pressure: 0mmHg to 299 mmHg
Pulse: 40 to 180/min.
Accuracy Pressure: ±3mmHg
Pulse: ± 5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Pressure Detection Capacitive pressure sensor
Memory 90 Measurements with date and time
Power Source 4 “AA” batteries 1.5V or AC/ DC adapter
(optional, 6V = 4W)
Battery Life Capacity of new alkaline batteries is approx. 1500
measurements
Operating temperature/
Humidity +10°C to +40°C
Maximum: 30 to 90% RH
Storage temperature/
Humidity -20°C to +60°C
Maximum: 10 to 95% RH
Console Weight Approximately 350g without batteries
Cuff Weight Approximately 240g
Outer Dimensions Approximat ely 131 (l) mm × 155 (w) mm × 84(h) mm
Cuff Dimensions Approximately 152 mm × 600 mm
(Cuff: arm circumference 22 to 42 cm)
Package Content Cuff, instruction manual, storage case, battery set,
guarantee card, blood pressure pass

= Type B

HEM-7000 main.book Page 26 Monday, June 19, 2006 8:53 AM