Troubleshooting

Trouble

Possible cause

Corrective action

 

 

 

 

Batteries are low (the orange

Replace the batteries with new

 

battery low indicator flashes).

ones. Refer to page 9.

The nebulisation rate is

 

 

The mesh is stained or clogged.

1. Clean the mesh by boiling. Refer

extremely low.

 

to page 16.

 

 

2. Replace the mesh cap with a new

 

 

one. Refer to page 11.

 

The batteries are low.

Replace the batteries with new

 

 

ones. Refer to page 9.

 

The capacity of rechargeable

Charge the batteries with a

The green power

batteries is low.

commercially available recharger.

The batteries are inserted in the

Re-insert the batteries in the correct

indicator does not turn

wrong direction.

direction. Refer to page 9.

on and the unit does not

The AC adapter is not correctly

Remove the AC adapter from the

nebulise.

 

connected to the main unit.

main unit and connect it correctly.

 

The medication container is not

Remove the medication container

 

installed correctly to the main unit.

from the main unit and install

 

 

correctly. Refer to page 12.

 

The vibrator is not immersed in the

Tilt the main unit in order to immerse

The green power

medication.

the vibrator in the medication.

indicator turns on

The medication container is not

Fill with medication. Refer to page

but the unit does not

filled with medication.

12.

nebulise.

The mesh is broken.

Replace with a new one. Refer to

 

 

 

page 11.