DISPLAY SYMBOLS

IRREGULAR HEARTBEAT SYMBOL ()

When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the measurement values.

An irregular heartbeat rhythm is defined as a rhythm that is more than 25% slower or 25% faster than the average rhythm detected while the monitor is measuring the systolic blood pressure and the diastolic blood pressure.

If the irregular heartbeat symbol ( ) displays with your measurement results, we recommend you consult your physician. Follow the directions of your doctor.

Normal Heartbeat

Pulse

Blood pressure

Irregular Heartbeat

Short Long

Pulse

Blood pressure

MOVEMENT ERROR SYMBOL ( )

The Movement Error Symbol ( ) is displayed if you move your body during the measurement. Please remove the arm cuff, wait 2-3 minutes, reapply the arm cuff, then take another measurement.

AVERAGE VALUE SYMBOL ( )

The Average Value Symbol ( ) is displayed when you press the MEMORY button. The most recent average reading appears on the display screen.

10

Page 10
Image 10
Omron Healthcare HEM-711AC Display Symbols, Irregular Heartbeat Symbol, Movement Error Symbol, Average Value Symbol