TROUBLESHOOTING GUIDE

PROBLEM

CAUSE

SOLUTION

 

 

 

 

The AC power cord is not

Turn the power switch off.

 

Plug the power plug into an

 

plugged into an electrical

No power on unit when

outlet.

electrical outlet. Turn the device

The rechargeable battery is

on. Make sure the battery is

the power switch is on.

correctly connected to the

 

not connected correctly,

 

compressor. Charge the

 

not sufficiently charged, or

 

battery. Replace the battery with

 

the battery has expired.

 

a new one.

 

 

 

No medication in the

 

 

medication cup.

Add the correct amount of

 

Too much or too little

prescribed medication to the

 

medication in the

medication cup.

 

medication cup.

 

 

 

 

 

The baffle is not placed in

Make sure the baffle is correctly

 

the medication cup or

positioned over the nozzle in the

 

incorrectly positioned.

medication cup.

 

 

 

 

The nebulizer kit is not

Make sure the nebulizer kit is

 

correctly assembled and the

 

correctly assembled.

inhalation accessory is correctly

 

 

attached.

 

The nozzle is blocked.

Clean and disinfect the nebulizer

No nebulization or low

kit to remove the blockage.

 

 

 

nebulization rate when

The nebulizer kit is tilted

Hold the nebulizer kit correctly.

the power is on.

Do not tilt the nebulizer kit so

 

at an incorrect angle.

the angle of the kit is greater

 

 

than 45 degrees.

 

 

 

 

The air tube is incorrectly

Make sure the air tube is correctly

 

attached to the compressor and

 

attached.

 

the nebulizer kit.

 

 

 

 

 

 

The air tube is folded

Make sure the air tube is not

 

folded, kinked or bent. Inspect

 

or damaged.

 

the air tube for any damage.

 

The air tube is blocked.

 

Replace the air tube if damaged.

 

 

 

 

 

 

The air filter is dirty.

Replace the air filter with a new

 

clean air filter.

 

 

 

 

 

The device is

The air filter cover is

Attach the air filter cover

correctly. Make sure the air filter

abnormally loud.

incorrectly attached.

cover is not blocked.

 

 

 

The compressor

Do not cover the compressor

 

with any type of cover during

 

is covered.

The device is very hot.

use. Turn the device off. Wait 40

The device has been used

 

minutes before using the device

 

for longer than 20 minutes.

 

again.

 

 

 

 

 