Panasonic EW3037 operating instructions Specifications

Models: EW3037

1 14
Download 14 pages 20.92 Kb
Page 14
Image 14

Symptom

Possible cause

 

 

The systolic pressure

• The hand with the cuff was held too low and

value or diastolic

not at heart level. (See page 12)

pressure value is high.

• The cuff was not attached properly. (See

 

page 12.)

 

• You moved your hand or body or you spoke

 

during measurement. (See page 12.)

 

 

The systolic pressure

• The hand with the cuff was held too high and

value or diastolic

not at heart level. (See page 12.)

pressure value is low.

• You moved your hand or body or you spoke

 

during measurement. (See page 12.)

 

 

The measured values

• Your posture was not the same for all

differ significantly.

measurements, or measurements were

 

taken without sufficient time in between.

 

Always wait approximately 5 minutes before

 

taking another reading, and above all, relax.

 

 

The measured value

• Blood pressure varies in response to

differs from that

minute changes in your mental state, such

measured by the doctor.

as your reaction to your doctor taking the

The measured value is

measurement. One way to check this is to

different for each

bring the unit to your doctor’s office and

measurement.

directly compare the readings taken by

 

your physician.

 

• Relax for at least 5 minutes and then take

 

another measurement.

 

 

Varying sounds and

• Changes in operating sounds and

pressurization speeds

pressurization speeds are normal and occur

occur during inflation of

due to the fact that measurements are

the cuff.

being made as the cuff is being inflated.

 

 

The measured value

• For persons with certain circulation problems,

obtained is different

there may be substantial differences in

from that measured at

blood pressure values measured at the

the upper arm.

upper arm versus at the wrist.

 

 

If the unit still appears to provide unusual or erroneous readings, consult your physician. If the unit does not appear to be functioning properly, contact Panasonic at 1-800-338-0552.

Specifications

Method of measurement:

Oscillometric System

 

 

Display:

Digital LCD

 

 

Measuring range:

Blood pressure: 0 - 280 mmHg

 

Pulse rate: 30 - 160 pulse/min.

 

 

Measurement accuracy:

Pressure: ±3 mmHg

 

Pulse rate: ±5%

 

 

Power source:

Two 1.5-volt alkaline (AAA/LR03) batteries

 

 

Operating environment:

10°C to 40°C (50°F to 104°F)

 

30% to 85% RH

 

 

Storage environment:

-20°C to 60°C (-4°F to 140°F)

 

10% to 95% RH

 

 

Wrist circumference

12.5 to 22.0 cm (5" to 8-3/4")

measurement range:

 

 

 

Dimensions (H#W#D):

54.0 #83. 0 #26.0 mm

 

(2-1/8" #2-1/4" #1-1/32")

 

 

Weight:

90 g (3.2 oz.) not including batteries

 

 

Specifications are subject to change without notice.

25

26

Page 14
Image 14
Panasonic EW3037 operating instructions Specifications

EW3037 specifications

The Panasonic EW3037 is a state-of-the-art digital thermometer designed to provide accurate and reliable temperature readings for users seeking to monitor their health. With its sleek and ergonomic design, the EW3037 is not just a tool but a companion for health-conscious individuals and families alike.

One of the standout features of the Panasonic EW3037 is its advanced infrared technology. This allows the thermometer to measure body temperature quickly and efficiently, with results available in just seconds. The infrared sensor captures the heat emitted from the body, ensuring precise readings without the discomfort of traditional thermometers. This makes it particularly suitable for use on children and infants, where quick readings can be crucial.

The EW3037 also boasts a wide measurement range, typically spanning from 32°C to 42.9°C. This capability means it can accommodate various needs, from fever detection to monitoring overall body temperature during different stages of health. The thermometer is designed to be user-friendly; it features a large digital display that makes it easy to read results, even in low-light conditions.

Another notable characteristic of the Panasonic EW3037 is its memory function. The device can store multiple previous measurements, enabling users to track temperature fluctuations over time. This is particularly helpful for individuals managing chronic conditions or those who need to monitor their health closely, as it allows for a more comprehensive understanding of their body temperature trends.

The EW3037 also emphasizes hygiene and safety. The thermometer is designed for easy cleaning, minimizing the risk of cross-contamination, which is essential in households with multiple users. Additionally, the device is lightweight and portable, making it easy to carry during travel or outings.

In terms of power efficiency, the Panasonic EW3037 typically utilizes easy-to-find batteries, ensuring that users can quickly replace them as needed. This thoughtful design consideration enhances its usability, as users are less likely to encounter interruptions in their health monitoring routine.

With its combination of advanced technology, user-friendly design, and practical features, the Panasonic EW3037 digital thermometer stands out as an essential tool for modern health management. Whether for personal use or a family's needs, it delivers a reliable means of keeping track of one of our most vital health indicators – body temperature – with ease and precision.