25)Alter the field to collect a standard powder measurement. See Figure 3.

26)Press [Esc] twice to return to the main APD menu.

27)Choose DATA COLLECTION (D).

28)Choose IDENTIFY MEASUREMENT (I).

29)Enter the name of the identify program that was just created. Press [Enter]. Press F5 to see a list of available identify programs if necessary.

30)Enter the sample identification and a filename where the data will be stored in the appropriate fields.

31)Check for the following:

ƒSample should be on the stage

ƒInstrument door should be closed.

ƒPower should be set to desired levels.

ƒShutter should be open.

ƒSee instructions for turning on generator if it is not on or at the proper power.

32)Push F1 to start the measurement.

33)To view the measurement as it is being collected, select DATA COLLECTION (D) and then DISPLAY SCAN (D).