6. SPECIFICATIONS

Electrical Requirements

115VAC±10￿, 60Hz / 230VAC±10￿, 50Hz

 

 

 

 

 

 

Rated Current Input

3.5A (115VAC/60Hz); 2.0A (230VAC/50Hz)

 

 

 

 

 

 

Sound Level

45dB(A) Average

 

 

 

 

 

 

 

 

Operating Altitude

Do not use at altitude over 4000 meters.

 

 

 

 

 

 

Oxygen Concentration

￿90￿at 1~5LPM

 

 

 

* Stated performance maximum achieved after 60 min. of run time.

Maximum Oxygen

6±0.5psi. (40±4 kPa.)

 

 

Delivery Pressure:

 

 

 

 

 

 

 

 

 

 

 

Optional Oxygen Sensor

 

 

 

 

 

 

 

Oxygen Purity

LED Indicator

 

 

 

 

 

 

 

 

 

Over 85￿(±3￿)

GREEN Indicator light

 

 

 

 

 

 

 

 

 

70￿(±3￿) to 85￿(±3￿)

YELLOW Indicator light

 

 

 

 

 

 

 

 

 

Below 70￿(±3￿)

RED Indicator light

 

 

 

 

 

 

 

 

 

 

 

 

Flow Range / Maximum

0-5 LPM / 5 LPM

 

 

Recommended Flow

 

 

 

 

 

 

 

 

 

 

Average Power

320W

 

 

Consumption

 

 

 

 

 

 

 

 

 

 

Pressure Relief Activation

250kPa±10￿

 

 

 

 

 

Filters:

Cabinet, Bacteria and Compressor Inlet

 

 

 

Alarms – (Visual/Audible)

High temperature w/ compressor – power shutdown

 

High / Low pressure alarm

 

 

 

Power loss alarm

 

 

 

 

 

Dimensions

68cm(26.8”) High × 45cm(17.7”) Wide × 33cm(13”) Deep

 

 

 

 

Circuit Breaker

(115VAC/60Hz)

 

(230VAC/50Hz)

& Fuse

Circuit Breaker: 8A

 

Circuit Breaker: 4A

Weight

28.2Kg (62lb)

 

 

 

 

 

 

Shipping Weight

31.8Kg (70lb)

 

 

 

 

 

 

Operating Ambient

10￿ ~ 39￿

 

 

Temperature

 

 

 

 

 

 

 

 

 

 

 

7

Page 8
Image 8
Pioneer Q150 owner manual Specifications