Medical Package System
BEFORE START UP
1.Make sure all safety warnings, labels and instructions have been read and understood before continuing.
2.Remove any shipping materials, brackets, etc.
3.Confirm that the electric power source and ground have been firmly connected.
4.Be sure all pressure connections are tight.
5.Check to be certain all safety relief valves, etc., are correctly installed.
6.Check that all fuses, circuit breakers, etc., are the proper size.
7.Make sure the inlet filter is properly installed.
8.Confirm that the drain valve is closed.
9.Visually check the rotation of the compressor pump. If the rotation is
incorrect, have a qualified electrician correct the motor wiring.
1.Follow all the procedures under
“Before
2.Switch the electric source breaker on.
3.Both dryers should be plugged in and set at the on position.
4.Dewpoint & CO Monitors should be on and the sampling valves open. Calibrate Dewpoint and CO per enclosed manufacturer’s instructions.
5.Ensure water lines are properly installed to
IMPORTANT: This is not a standard option. Check to see if water-
cooled aftercoolers were provided.
6.Compressor unit isolation valves are preset at the factory in the open position except for tank
7.Dryer and filter isolation valves: Valves should be open to one dryer and one filter bank.
8.Close valve leading to the Medical Air System from receiver on the compressor unit.
a.Turn both fusible/breaker disconnects to the on position. Low pressure alarm will sound. Silence the alarm by pushing alarm acknowledge button.
b.Jog each compressor in the manual position on the selector switch to check for rotation. (Clockwise if
Medical
Air Schematic
Specifications
Medical Package Duplex – Model MPD
MODEL | HP | PHASE | SCFM | SCFM | RPM | VOLTAGE | FULL LOAD AMP/ | GALLON | DIMENSION | SHIPPING |
@100 PSIG | @ 50 PSIG | MOTOR | TANK | LxWxH | WT. (Lbs.) | |||||
MPD0508 | 5(2) | 3 | 33.4 | 36.2 | 870 | 208/230/460 | 17.4 / 16.5 / 8.2 | 120 | 64 x 65 x 76 | 1288 |
|
|
|
|
|
|
|
|
|
|
|
MPD0758 | 7.5(2) | 3 | 52.2 | 55.1 | 840 | 208/230/460 | 23.4 / 22 / 11 | 120 | 64 x 65 x 76 | 1340 |
|
|
|
|
|
|
|
|
|
|
|
MPD1008 | 10(2) | 3 | 70 | 76.3 | 1070 | 208/230/460 | 29 / 33 / 16.5 | 200 | 64 x 65 x 80 | 1510 |
|
|
|
|
|
|
|
|
|
|
|
MPD1508 | 15(2) | 3 | 108.4 | 114.1 | 1250 | 208/230/460 | 51 / 48 / 24 | 200 | 64 x 65 x 80 | 1630 |
|
|
|
|
|
|
|
|
|
|
|
Medical Package Triplex – Model MPT
MODEL | HP | PHASE | SCFM | SCFM | RPM | VOLTAGE | FULL LOAD AMP/ | GALLON | DIMENSION | SHIPPING |
@100 PSIG | @ 50 PSIG | MOTOR | TANK | LxWxH | WT. (Lbs.) | |||||
MPT0508 | 5(3) | 3 | 50.1 | 54.7 | 870 | 208/230/460 | 17.4 / 16.5 / 8.2 | 200 | 96 x 65 x 80 | 1838 |
|
|
|
|
|
|
|
|
|
|
|
MPT0758 | 7.5(3) | 3 | 78.8 | 83.9 | 840 | 208/230/460 | 23.4 / 22 / 11 | 200 | 96 x 65 x 80 | 1980 |
|
|
|
|
|
|
|
|
|
|
|
MPT1008 | 10(3) | 3 | 105 | 111.5 | 1070 | 208/230/460 | 29 / 33 / 16.5 | 240 | 96 x 65 x 92 | 2160 |
|
|
|
|
|
|
|
|
|
|
|
MPT1508 | 15(3) | 3 | 162.6 | 168 | 1250 | 208/230/460 | 51 / 48 / 24 | 240 | 96 x 65 x 92 | 2280 |
|
|
|
|
|
|
|
|
|
|
|
Medical Package Quadplex – Model MPQ
| MODEL | HP | PHASE | SCFM | SCFM | RPM | VOLTAGE | FULL LOAD AMP/ | GALLON | DIMENSION | SHIPPING |
|
| @100 PSIG | @ 50 PSIG | MOTOR | TANK | LxWxH | WT. (Lbs.) |
| |||||
| MPQ0508 | 5(4) | 3 | 66.8 | 74.9 | 870 | 208/230/460 | 17.4 / 16.5 / 8.2 | 240 | 128 x 65 x 92 | 2388 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| MPQ0758 | 7.5(4) | 3 | 104.4 | 120.2 | 840 | 208/230/460 | 23.4 / 22 / 11 | 240 | 128 x 65 x 92 | 2490 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| MPQ1008 | 10(4) | 3 | 140 | 153 | 1070 | 208/230/460 | 29 / 33 / 16.5 | 240 | 128 x 65 x 92 | 2640 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| MPQ1508 | 15(4) | 3 | 216.8 | 221 | 1250 | 208/230/460 | 51 / 48 / 24 | 240 | 128 x 65 x 92 | 2735 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
4