Raytek MID

Measurement Specifications

Spectral Response:

8 to 14 microns

Optical Resolution:

2:1 or 10:1

Temperature Range:

-40 to 600°C (-40 to 1112°F); -25 to 600°C for

 

J-thermocouple output

System Accuracy:

±1% of reading or ±1°C, whichever is greater

 

@23°C±5°C (73°F±9°F). Thermocouple output

 

accuracy ±1% of reading or ±2.5°C, whichever is

 

greater @23°C±5°C

System Repeatability:

±0.5% of reading or ±0.5°C (1°F), whichever is greater

Temperature Coefficient:

0.15K per K or 0.15% per K, whichever is greater

Temperature Resolution:

0.3°C or 0.5°F

System Response Time:

150ms (95%)

Emissivity:

0.100 to 1.100 digitally adjustable increments of .001

Transmission:

0.100 to 1.100 digitally adjustable increments of .001

Signal Processing:

Peak Hold, Valley Hold, Variable Averaging Filter,

 

adjustable up to 998 seconds

Electrical Specifications

Outputs:

Scalable 4-20mA, 0-20mA, 0-5V,

 

J or K thermocouple

 

10mV/°C Head Ambient signal

Cable Length

1 m (3.2 ft) standard, 3m (9.8 ft.) optional

Output Impedance (T/C output)

20 ohm

Minimum Load Impedance (mV output)

100K ohms

Maximum Loop Impedance (mA output)

500 ohms with 24 VDC power supply

Current Draw:

100 mA

Power Supply

11-26 VDC

Nominal Optical Specifications

D:S is the optical resolution expressed as a ratio of the distance to the resolution spot divided by the diameter of the spot.

Optical resolution for the MID is 2:1 and 10:1. Nominal spot size based on 90% energy.

General Dimensions

Sensing Head

Electronics Housing

25

(1.0)

General Specifications

Environmental Rating

NEMA-4 (IP 65)

Ambient Temperature Range:

 

Sensing head

0 to 85°C (32 to 185°F)

With air cooling

-18 to 200°C (0 to 392°F)

Electronics housing

0 to 65°C (32 to 150°F)

Storage Temperature

-18 to 85°C (0 to 185°F)

Relative Humidity

10 to 95%, non-condensing

Construction:

 

Sensing head

Stainless steel

Electronics housing

Zinc, die-cast

Weight:

 

Sensing head (w/1 m cable)

50 g (1.75 oz)

Electronics housing

270 g (9.5 oz)

 

60

80

 

(3

(2

 

.2)

 

mm (in)

.

 

4)

 

Accessories / Options*

Each standard MID package includes a sensing head, one mounting nut,

1 m (3.2 ft) of cable, die-cast housing with premounted electronics, and an operator’s manual

Adjustable or fixed mounting bracket

Air purge jacket

Air cooling/purging system

Longer cable: 3m (9.8 ft.)*

Rs232 or RS485 communication*

*Must be specified at time of order

For Sales and Ordering Contact:

5710 Kenosha Street Richmond, Illinois 60071 USA Phone: +1 (815) 678-2211 FAX: +1 (815) 678-3961 Internet: www.watlow.com e-mail: info@watlow.com

Your Authorized W atlow Distributor Is:

For Service and

W arranty Contact:

Worldwide Headquarters

Raytek Corporation

Santa Cruz, CA USA

Tel: 1 831 458 1110

800 227 8074 (U.S. and Canada only) Fax: 1 831 458 1239 www.raytek.com

Raytek, the Raytek logo, and Thermalert are registered trademarks of Raytek Corporation.

©2001 Raytek Corporation, Printed in USA, 3/01 Rev. A Specifications subject to change without notice. 3-1022

RIC-MID-0401

Page 2
Image 2
RayTek MID manual Measurement Specifications, Electrical Specifications, Nominal Optical Specifications, General Dimensions