Specifications and Features

RayngerST

Specifications

 

Temperature Range

 

ST20 Pro

-32 to 400˚C (-25 to 750˚F)

ST30 Pro

-32 to 545˚C (-25 to 950˚F)

 

 

Accuracy

1% or ±1˚C (±2˚F), whichever is greater above 23˚C (73˚F)

@23˚C (73˚F) ambient ±2˚C (±3˚F) @ -18 to 23˚C (0 to 73˚F)

operating temperature ±2.5˚C (±4˚F) @ -26 to -18˚C (-15 to 0˚F)

±3˚C (±5˚F) @ -32 to -26˚C (-25 to -15˚F)

Repeatability

±0.5% or ±1˚C (±2˚F) whichever is greater

 

 

 

Response Time

0.5 second (95% of reading)

 

 

 

Ambient Operating Range

0 to 50˚C (32 to 120˚F)

 

 

 

Power

9V Alkaline or NiCd Battery

 

 

 

Weight

320g (11oz)

 

 

 

 

 

Features

 

ST20 Pro

ST30 Pro

Distance to Spot (D:S)

 

12:1

12:1

Emissivity

 

pre-set at 0.95

pre-set at 0.95

Laser Sighting

 

single point

offset 8 point circular

MAX Temperature Display

 

Display Hold (7 seconds)

 

LCD Backlight

 

Display Resolution

 

0.2°C (0.5°F)

0.2°C (0.5°F)

Hard Carrying Case

 

Options/Accessories

Nylon Holster

NIST Traceable Calibration Certificate

ST20 and ST30 ProPlus Target Spot Sizes D:S = 12:1 at Focus Point

For Sales and Ordering Contact:

5710 Kenosha Street Richmond, Illinois 60071 USA Phone: +1 (815) 678-2211 FAX: +1 (815) 678-3961 Internet: www.watlow.com e-mail: info@watlow.com

Your Authorized Watlow Distributor Is:

For Service and

Warranty Contact:

Worldwide Headquarters

Raytek Corporation

Santa Cruz, CA USA

Tel: 1 831 458 1110

800 866 5478 (U.S. and Canada only) Fax: 1 831 425 4561 www.raytek.com

Raytek, the Raytek logo, and Thermalert are registered trademarks of Raytek Corporation.

©2001 Raytek Corporation, Printed in USA, 3/01 Rev. A Specifications subject to change without notice. 3-1022

RIC-20-0401

Page 2
Image 2
RayTek ST30, ST20 manual RayngerST