Clinical Use Only
Disassembling and Reassembling the HumidAire 2i
WARNING
Beware of electrocution. Do not immerse the device or power cord in water. Always unplug the device before cleaning and be sure that it is dry before plugging back in.
1Turn the main power switch at the back of the device to OFF, and unplug the power cord from the power outlet.
2Allow 10 minutes for the tub plate and any excess water to cool.
3Remove the water chamber. To do this, lift the latch at the bottom of the water chamber and pull the water chamber away from the docking station.
4Discard any excess water by turning the water chamber upside down, allowing the water to drain out of the air outlet.
5Disassembling the Water Chamber
1. To open the side catches pull out from the lower edge.
Side catch
2. Separate the water chamber by pulling the two halves apart. Use your thumb to pull down on the front tab.
Front tab
3.Inside the chamber lid is the seal. Remove the seal by peeling it away from the inside of the lid.
Lid
Seal
4. Remove the
Base
5