MDF-U3386S
Features
EVOLUTIONARY DESIGN
•The newly developed refrigeration system and the freezer structure offer a quiet experimental environment.
•Achieves 22% higher energy conser- vation than previous model. (230V, 50Hz)
•The newly developed back cover combined with new aerodynamically designed and placed components in the refrigeration compartment provide superior air flow, making it possible to
drastically reduce | Compressors | |
the stress to the Front |
| Back |
freezer and |
| cover |
|
| |
contributing to |
|
|
excellent |
|
|
durability. | Rear | Air |
•Two independent and insulated inner doors prevent cold air leakage.
ACCESSIBILITY
• A new beak style inner door latch tightly closes the inner door against the freezer
frame. It also helps make opening and closing the door smoother.
•The condenser filter is situated at the bottom right side of the front panel to make filter removal and cleaning easier.
SAFETY
•Alarm lamp and buzzer offer secure warning of power failure or abnormal temperature increase.
•High and low temperature warning provides an audible and visual alarm
when the temperature deviates more than ±5˚C to ±20˚C (adjustable) from the set point.
• Alarm ring back function ensures that buzzer will resume operation, should alarm conditions continue after it is silenced.
•
•New rugged,
•Control panel with digital display for easy operation.
Performance Data
Pull-down characteristics
| 40 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
(˚C) | 20 |
|
|
| Central air temp. in cabinet |
|
| ||||||
0 |
|
|
| Ambient temp. 30˚C, |
| ||||||||
|
|
|
|
| |||||||||
Temperature |
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
| ||
| 2 | 4 | 6 | 8 | 10 | 12 | 14 | ||||||
| 0 |
Time (hour)
| 40 | |
(˚C) | 20 | |
0 | ||
Temperature | ||
| ||
|
0 | 2 | 4 | 6 | 8 | 10 | 12 | 14 |
|
|
| Time (hour) |
|
|
CO2 backup system operation characteristics
20 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| Ambient temp. and maintenance |
|
|
| |||||||||
(hour) |
| of |
|
| ||||||||||
15 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
Time |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
10 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
5 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 10 | 15 | 20 | 25 | 30 |
| 35 | |||||||
5 |
|
Ambient temperature (˚C)
Dimensions |
|
|
| ||
|
| 875 (34.4) |
| 816 (23.1) |
|
105 |
| 795 (31.3) | 48 | 750 (29.5) | 18 |
(4.1) | 140 | 600 (23.6) | (1.9) | 490 (19.3) | (0.7) |
| (5.5) |
| 130 (5.1) |
|
|
|
|
|
|
|
11 | (44.9) |
| (55.1) |
|
|
|
|
| |
(0.4) | 1140 |
| 1400 | (72.8) |
| 34 | |||
(58.5) | 38 | (1.3) |
| 1850 |
|
| |||
|
|
|
| |
1485 | (1.5) |
| 218 (8.6) |
|
(5.1) |
|
|
| |
| 130 |
| 450 (17.7) |
|
|
|
|
|
[Unit : mm (inch)]
Specifications
Model | |
|
|
Temperature range | |
|
|
Maximum cooling | |
performance | (Ambient temp. 30˚C) |
|
|
Exterior dimensions | 750 x 875 x 1850mm |
(W x D x H) | (29.5” x 34.4” x 72.8”) |
|
|
Interior dimensions | 490 x 600 x 1140mm |
(W x D x H) | (19.3” x 23.6” x 44.9”) |
|
|
Net weight | Approx. 255 kg (562 lbs.) |
|
|
Effective capacity | 333 L (11.8 cu.ft.) |
|
|
Shelf | Stainless steel, Adjustable, |
| 3 shelves, Max. load: 50kg |
| (110 lbs.)/shelf |
|
|
Access port | 17mm diameter, 3 locations |
| (back, bottom left/right corner) |
|
|
Compressor | Hermetic type, Output: 450W |
| (high stage side), 750W (low |
| stage side) |
|
|
Refrigerant | HFC refrigerant |
|
|
Alarm | High/low temperature, Power |
| failure, Filter check, Battery |
| check, Part replacement |
| notification |
|
|
Remote alarm contact | Allowable contact capacity: |
| DC 30V, 2A |
Accessories | 1 set of keys, 1 scraper |
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30°C with no load). In order to use the freezer at a stable temperature for a long time, it is recommended that the temperature be set to at least 5°C higher than the indicated lowest temperature.
In addition, depending on the usage conditions, it may not be possible to reach the indicated lowest temperature.
Caution:
SANYO guarantees the product under certain warranty conditions. SANYO in no way shall be responsible for any loss of content or damage to content.
•Appearance and specifications are subject to change without notice.
Options
Rack Capacity
| For |
| ||
|
|
| 6 | |
|
|
| 6 | |
| For |
| ||
|
|
| 12 | |
Inventory rack | Inventory rack |
| ||
| ||||
| Strip chart recorder |
| ||
|
| |||
| Recorder mounting |
| ||
| bracket |
| ||
Circular recorder |
| |||
(for |
|
SANYO DAQ (data acquisition) System | LCO2 backup system |
| |
Ethernet (LAN) Interface |
|
or RS232C/RS485 Interface |
|
SANYO Electric Co., Ltd., Biomedical Division, Gunma is certified for:
Quality management system: ISO9001 / Medical devices quality management system: ISO13485 / Environmental management system: ISO14001
Distributed by:
SANYO Electric Co., Ltd.
Biomedical Division
http://biomedical.sanyo.com/
©2009 SANYO Printed in Japan 2009.9 MA