Failure Mode Comparison

Twin Guard Seriesvs. Conventional Cascade Freezer

 

 

Conventional Freezer,

 

Event

Conventional Freezer No Back-Up

with Back-Up System

SANYO Twin Guard Series

 

 

 

 

Fan Failure

Freezer Fails

 

 

High Stage Compressor Failure

Freezer Fails

 

 

Low Stage Compressor Failure

Freezer Fails

Freezer fails, CO2 or LN2 back-up system offers short term protection until contents can be removed and repairs initiated

If one fan fails the second fan automatically maintains energy exchange

No high or low stage used. Two refrig- eration systems, each with a separate compressor, operate independently in overlapping cycles during normal op- eration. If one compressor fails internal temperature is maintained indefinitely at -65°C with backup compressor. Optional LN2 or CO2 back-up system offers additional protection

 

20

 

(°C)

0

Pull-up while a power failure

 

temp.

-20

 

 

 

Chamber

-40

 

 

Single compressor operation

 

 

 

-60

(kept at approx. -65°C)

 

-80

 

Chamber temp. (°C)

40

20

0

-20

-40

-60

-80

Central air temp. in cabinet Ambient temp. 30°C, No-Load

-100

Normal cycle operation (-80°C setting)

0 1 2 3 4 5 6 7 8 9 10 11 12

(hour)

-100

0 1 2 3 4 5 6 7 8 9 10 11 12

(hour)

Performance Chart, power-failure situation Normal operation and Twin Guard Series™ backup.

Pull-up characteristics during power failure Pull-down charateristics

Reduced Power Consumption

The SANYO Twin Guard SeriesFreezer can be set for Normal or EcoModeoperation, depending on ambient tem- perature and load. EcoModeis recom- mended for 90-95% of applications. Although both refrigeration systems are completely independent, EcoModeestablishes an overlapping cycle to significantly reduce energy consump- tion while optimizing interior uniformity from top-to-bottom and front-to-back for protection of high value materials.

Power Consumption Benefits

Model Number

MDF-U500VXC

MDF-U700VXC

 

 

 

 

 

Ambient Temperature

15°C

23°C

15°C

23°C

 

 

 

 

 

Normal Savings

-9.7%

-7.8%

-16.7%

-17.5%

 

 

 

 

 

EcoModeSavings

-16.8%

-14.1%

-18.5%

-18.3%

 

 

 

 

 