DAILY INSPECTION INSTRUCTIONS
A shielded outlet port cover mechanism permits a secure attachment of source guide tubes (projection sheaths) without an elevation of radiation levels. The outlet port mechanism will not allow the source assembly to be projected outside of the exposure device unless a source guide tube has been properly attached to the outlet port. A
1Pull and rotate the
3o'clock position) in a clockwise direction. This action moves the protective cover away from the outlet port to permit attachment of a source guide tube.
2A
3Rotate the
The design permits attachment of the source guide tubes without an elevation of radiation intensity at the outlet port, minimizing the radiation exposure to the radiographer's hands. The sealed source assembly cannot be exposed out of the exposure device unless a positive connection of the source guide to the device has been accomplished.
The exposure device's locking mechanism provides three main safety features for the radiographer during radiographic operations.
First, the locking mechanism requires a deliberate action to enable exposure of the source assembly. This is accomplished by coupling the control cable connector to the source assembly connector, coupling the remote control connector to the exposure device locking mechanism and then pushing a lock slide from the SECURED position into the EXPOSE position. The lock slide must be manually pushed into the EXPOSE position prior to each radiographic exposure to enable operation of the remote control.
Source assembly
Sealed source
|
| Control cable | Exposure device | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Remote | Locking | Shield | Outlet port | Source |
control crank | mechanism |
|
| guide tubes |
Figure 3.2 Source assembly in fully shielded position of radiographic exposure device
3.3