Appendix A: Specifications
18 CPS250 User Manual
Table 4: Environmental Characteristics
Storage Temperature –10_ C to 60_ C, 80% RH
Operating Temperature 10_ C to 40_ C, 75% RH
Table 5: Operational Characteristics
Outputs 0 to 20 VDC (two), 5 VDC
Voltage (5 V) 5.0 ±0.1 VDC at 2.0 A maximum
Voltage (0–20 V) 0–20 VDC at 0.5 A maximum
Load Regulation (5 V) 0.1% +5 mV
Load Regulation (0–20 V) 0.01% +3 mV
Ripple/Noise 2 m VRMS, 5 Hz–1 MHz
Tracking Error ±0.2% ±20 mV
Indicator Analog type front panel meter
Meter Indicators 0–25 VDC ±2.5% of full scale
0–600 mA ±2.5% of full scale
Table 6: Electrical Characteristics
Line Voltage Range 90 to 110, 108 to 132, 198 to 242, and 216 to 250 VAC at
50–60 Hz
Power Consumption 175 VA, 160 W maximum