Performance Tests
TDS 520A, 524A, 540A, & 544A Performance Verification
4.
Confirm input channels are within limits for DC accuracy at maximumoffset and position: Do the following substeps — test CH 1 first,
skippingsubstep a since CH 1 is already selected from step 3.a.
Select an unchecked channel:Press WAVEFORM OFF to remove the channel just confirmed
from the display.
Press the front-panel button that corresponds to the channel you
are to confirm.
Set the generator output to 0 V.Move the test hookup to the channel you select.
b.
Turn on the measurement Mean for the channel:Press MEASURE, then press the main-menu button Select
Measrmnt for
CHx.
Press the side menu button more until the menu label Mean
appears in the side menu (its icon is shown at the left). Press the
side-menu button Mean.
Press CLEAR MENU.
Follow these rules to match this procedure to the model of the oscillo-
scope under test:
Models TDS 540A, 544A Only—Use Table 1-3 to test CH 1—CH 4;
ignore Table 1-4 AUX 1 & AUX 2 settings and limits.
Model TDS 520A, 524A Only—Use Table 1-3 to test CH 1 and CH 2
only; use Table 1-4 to test AUX 1 and AUX 2 only.
c.
Set its vertical scale: Set the vertical SCALE to one of the settings
listed in Table 1-3 (and Table 1-4 for the TDS 520A and 524A) that is
not yet checked. (Start with the first setting listed.)
Table1-3: DC Accuracy: CH 1–CH 4
Scale
Setting Position
Setting
(Divs)
Offset
Setting Generator
Setting Accuracy
Limits
5mV –5 +1 V +1.040 V +1.0355 V to +1.0445 V
+5 –1 V –1.040 V –1.0355 V to –1.0445 V
200 mV –5 +10 V +11.6 V +11.525 V to +11.675 V
+5 –10 V –11.6 V –11.525 V to –11.675 V
1V –5 +100 V +108 V +107.450 V to +108.550 V
+5 –100 V –108 V –107.450 V to –108.550 V