Appendix E: Performance Verification
VX4801 Programmable Digital I/O Module A–25
NOTE. If you are using National Instruments NI-488.2 software you may wish to
select the buffer 1 mode to allow more comfortable viewing of the ASCII
response. Just type buffer 1.
1. To verify the system configuration, send the TABLE command to the Slot 0
Resource Manager and confirm the responses shown in Table 6. Your
configuration may not be identical, but the responses should be similar.
Table 6: VXIbus System Configuration
Command to Type Response to Verify
("(#
("&(*$  
("/,- -!")%
(",$  
    )+-   '   
  . .
    )+-   '   
     

    )+-   '   
     

2. Verify the VX4801 VXIbus interrupt capability with the following steps:
NOTE. Make sure your Slot 0 controller and the VX4801 under-test are set to the
same interrupt level (see User Manual for location of interrupt setting). Also, if
you are using National Instruments NI-488.2 software, make sure Auto Serial
Polling is disabled to prevent the SRQ from being reset prior to a visual check.
a. Enable the generation of VXI Request True interrupt and force a VXIbus
interrupt with an error condition with the following commands:
("&(*$  
("/,- 0!%
(Enable VXI Request True interrupt)
("/,- .0(
(Observe: VX4801 ERR light is on)