EDIT Menu

2)Set the Cursor Point setting to Exclude.

3)Apply smoothing. However, there are waveforms for which smoothing may not be effective.

￿Increasing the Size of the Area. Data that is identical across the operation can be acquired by increasing the size of the area. (See Figure 3 -60.) Smooth data is acquired when this area is shifted, as shown in the right of the figure.

Original Data

Data Shifted to Right by 0.5

Figure 3 60: Increasing the Size of the Area

This is due to the fact that when using linear interpolation to derive data values between points, the value of the data does not change when interpolating between points with the same value. The value not changing is equivalent to not performing a shift. However, this error can be made inconspicuous by the insertion of an appropriate anti-aliasing filter.

Multiple Copy. The Multiple Copy... item appears in the menu when two or more waveforms are being edited simultaneously. It is used to copy the section of the waveform between the two vertical bar cursors in one editing area to the waveform between the vertical bar cursors in another editing area, at the interval specified with Interval. The following diagram shows the menu configuration for the Multiple Copy... item.

Source

Interval

Operation ￿More 4 of 4￿Multiple Copy...Go Back

Execute

See Page 3 -15 ªOpening and Selecting Editing Areasº for further information on designating multiple editing areas.

In this example, we will copy the waveform between the vertical bar cursors in editing area 1 to editing area 2. We will start from the point at which only one waveform has been designated in the waveform editor.

AWG2021 User Manual

3 79