SETUP Menu
Setting the amplitude and offset determines Max and Min values shown to the left of the waveform. In the example shown in Figure 3
Max: 3.500V/50W
Min:
Setting Amplitude 1. Select Amplitude from the bottom menu. The amplitude icon will be highlighted on the screen.
2.Select the channel from the side menu.
3.Use the numeric keys or the general purpose knob to set the amplitude. The output amplitude can be set to any value between 0.05 V and 5 V in minimum increments of 1 mV.
NOTE. The amplitude set gives the top and bottom voltage values for the waveform editor
Setting Offset 1. Select Offset from the bottom menu. The offset icon will be highlighted on the screen.
2.Select the channel from the side menu.
3.Use the numeric keys or the general purpose knob to set the desired offset. The offset may be set to any value between ±2.5 V and +2.5 V in minimum increments of 5 mV.
Linking the CH1 Waveform When Option 02 is installed, it is possible to link the amplitude and offset of and Amplitude or Offset CH1 with the corresponding values for CH2. When such linkages are set,
changes to CH1 values will also be applied to the CH2.
Use the following procedure to link CH2 amplitude to CH1 amplitude. In such cases, CH1 and CH2 have their own files one for each.
3 226 | AWG2021 User Manual |