Appendix C: Performance Verification
b.Hook up oscilloscope:
Connect the oscilloscope probe to the CH1 vertical input.
Connect the probe
c.Set oscilloscope controls:
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . CH1
Scale . . . . . . . . . . . . . . . . . . . . . . . . 1 V/div.
Input Impedance . . . . . . . . . . . . . . 1 MW
Horizontal
Sweep . . . . . . . . . . . . . . . . . . . . . . . Adjust as needed
Trigger
Mode . . . . . . . . . . . . . . . . . . . . . . . Auto
2.Create the AWG2021 waveform file, select waveform file, and set AWG2021 controls:
a.Initialize AWG2021 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. If Option 02 is installed (Option 02 adds the CH2 output channel), Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the DIGI_OUT.WFM file.
Push ENTER to select the file.
c.Set AWG2021 controls:
Push MODECont
3.Check the CH1 digital data output signals:
a.Check the signal levels:
Contact the oscilloscope probe to the pins on 2 ⋅ 13 header (see Figure
Data signals
C48 | AWG2021 User Manual |