Appendix C: Performance Verification
Floating Point Processor Check (Option 09)
This procedure checks the AWG2021 floating point processor. This check requires that the AWG2021 has Option 09 installed.
Equipment Required | None. |
|
|
Prerequisites | The AWG2021 meets the prerequisites listed on page C5. |
|
|
1.Check that floating point processor test in internal diagnostics passes:
a.Run the AWG2021 internal diagnostics: Push the AWG2021 ON/STBY switch two times so that the AWG2021 runs the
b.Check the FPP test results: When the AWG2021 finishes the FPP test, check that the test result is Pass.
This completes the performance tests for the AWG2021.
C50 | AWG2021 User Manual |