DARK_MODE 5 AUT O OFF,
AUTO,
CMND
Dark calibration mode. OFF
disables it; AUTO does it
periodically; CMND does it only
when commanded. Enclose
value in double quotes (“) when
setting from the RS-232
interface.
DARK_ENABLE 1, 3, 4 ON,
OFF 4
OFF, ON ON enables PMT/UV dark
calibration; OFF disables it.
DARK_FREQ Minutes 30,
720 3
0.1–1440 Dark calibration period.
DARK_PRE_DWELL Seconds 10 1–60 Dwell time after closing dark
shutter or turning off lamp or
selecting preamp range.
DARK_POST_DWELL Seconds 10,
30 3
1–180 Dwell time after opening dark
shutter or turning on lamp.
DARK_SAMPLES Samples 5 1–10 Number of dark samples to
average.
DARK_FSIZE Samples 2 1–100 Dark offset moving average filter
size.
DARK_LIMIT mV 200,
400 3
0–1000 Maximum dark offset allowed.
SO2_TARG_ZERO1 Conc 0 -100–999.99 Target SO2 concentration during
zero calibration of range 1.
SO2_SPAN1 Conc 400,
4000 3
0.01–9999.99 Target SO2 concentration during
span calibration of range 1.
SO2_SLOPE1 PPB/mV,
PPM/mV 3
1 0.25– 4 SO2 slope for range 1.
SO2_OFFSET1 mV 0 -1500–1500 SO2 offset for range 1.
SO2_TARG_ZERO2 Conc 0 -100–999.99 Target SO2 concentration during
zero calibration of range 2.
SO2_SPAN2 Conc 400,
4000 3
0.01–9999.99 Target SO2 concentration during
span calibration of range 2.
SO2_SLOPE2 PPB/mV,
PPM/mV 3
1 0.25– 4 SO2 slope for range 2.
SO2_OFFSET2 mV 0 -1500–1500 SO2 offset for range 2.
CO2_TARG_SPAN_CONC 11 % 12 0.1–1000 Target CO2 concentration during
span calibration.
CO2_SLOPE 11 1 0.5–5 CO2 slope.
CO2_OFFSET 11 % 0 -10–10 CO2 offset.
O2_TARG_SPAN_CONC 10 % 20.95 0.1–100 Target O2 concentration during
span calibration.
O2_SLOPE 10 1 0.5–2 O2 slope.
O2_OFFSET 10 % 0 -10–10 O2 offset.
RANGE_MODE SNGL SNGL,
DUAL,
AUTO,
Range control mode. Enclose
value in double quotes (“) when
setting from the RS-232
interface.