TELEDYNE INSTRUMENTS
Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation
05928 Rev F.0B A-17
DCN 5063 PRINTED DOCUMENTS ARE UNCONTROLLED
19200,
38400,
57600,
115200
MODEM_INIT2 “AT Y0 &D0
&H0 &I0 S0=2
&B0 &N6 &M0
E0 Q1 &W0”
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 COM2 modem
initialization string. Sent verbatim
plus carriage return to modem on
power up or manually. Enclose
value in double quotes (“) when
setting from the RS-232
interface.
RS232_PASS Password 940331 0–999999 RS-232 log on password.
MACHINE_ID ID 100 0–9999 Unique ID number for instrument.
COMMAND_PROMPT “Cmd> ” Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 interface command
prompt. Displayed only if enabled
with RS232_MODE variable.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
TEST_CHAN_ID NONE NONE,
PMT
READING,
UV READING,
VACUUM
PRESSURE3,
SAMPLE
PRESSURE,
SAMPLE
FLOW,
RCELL TEMP,
O2 CELL
TEMP 10,
CHASSIS
TEMP,
IZS TEMP 1,
PMT TEMP,
HVPS
VOLTAGE
Diagnostic analog output ID.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
REMOTE_CAL_MODE LOW LOW,
HIGH,
CO2 11,
O2 10
Range to calibrate during
contact-closure and Hessen
calibration. Enclose value in
double quotes (“) when setting
from the RS-232 interface.
PASS_ENABLE OFF OFF, ON ON enables passwords; OFF
disables them.
STABIL_FREQ 99 Seconds 10 1–300 Stability measurement sampling
period.
STABIL_SAMPLES 99 Samples 25 2–40 Number of samples in
concentration stability reading.
RCELL_CYCLE Seconds 2 0.5–30 Reaction cell temperature control
cycle period.