Table A-3: M100EU Test Functions, Revision F.0B
Name 1 Message Text Description
RANGE RANGE= 500.0 PPB 3
SO2 RNG=500.0 PPB 3, 10, 11
D/A range in single or auto-range modes.
RANGE1 RANGE1=50 0.0 PPB 3
SO2 RN1=500.0 PPB 3, 10, 11
D/A #1 range in independent range mode.
RANGE2 RANGE2=50 0.0 PPB 3
SO2 RN2=500.0 PPB 3, 10, 11
D/A #2 range in independent range mode.
CO2RANGE CO2 RNG=100 PCT 11 CO 2 range.
O2RANGE O2 RNG=100 WT% 10 O
2 range.
STABILITY ST ABIL=0.0 PPB 3
SO2 STB=0.0 PPB 3, 10
O2 STB=0.0 WT% 10
CO2 STB=0.0 PCT 11
Concentration stability #1. Configure with
SETUP-MORE-STBL.
STABILITY2 6 ST ABIL2=0.0 PPB 3
SO2 STB2=0.0 PPB 3, 10
O2 STB2=0.0 WT% 10
CO2 STB2=0.0 PCT 11
Concentration stability #2. Configure with
SETUP-MORE-STBL.
RESPONSE 2 RSP= 1.11(0.00) SEC Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
VACUUM 5 VAC=9.1 IN-HG-A Vacuum pressure.
SAMPPRESS PRES=29.9 IN-HG-A Sample pressure.
SAMPFLOW SAMP FL=700 CC/M Sample flow rate.
PMTDET PMT=762.5 MV Raw PMT reading.
NORMPMTDET NORM PMT=742.9 MV PMT reading normalized for temperature,
pressure, auto-zero offset, but not range.
UVDET UV LAMP=3457.6 MV UV lamp reading.
STABILITYUV 6 UV ST B=5.607 MV UV lamp stability reading.
LAMPRATIO LAMP RATIO=100.0 % UV lamp ratio of current reading divided
by calibrated reading.
STRAYLIGHT STR. LGT=0.1 PPB Stray light offset.
DARKPMT DRK PMT=19.6 MV PMT dark offset.
DARKLAMP DRK LMP=42.4 MV UV lamp dark offset.
SLOPE SLOPE=1.061 Slope for current range, computed during
zero/span calibration.
OFFSET OF FSET=250.0 MV Offset for current range, computed during
zero/span calibration.
CO2SLOPE 11 CO2 SLOPE=1.0000 CO2 slope, computed during zero/span
calibration.
CO2OFFSET 11 CO2 OFFSET=0.00 % CO2 offset, computed d uring zero/span
calibration.
O2SLOPE 10 O2 SLOPE=0.980 O2 slope, computed during zero/span
calibration.
O2OFFSET 10 O2 OFFSET=1.79 % O2 offset, computed during zero/span