Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation
05928 Rev F.0B A-23
DCN 5063 PRINTED DOCUMENTS ARE UNCONTROLLED
calibration.
HVPS HVPS=650 VOLTS High voltage power supply output.
RCELLDUTY RCELL ON=0.00 SEC Reaction cell temperature control duty
cycle.
RCELLTEMP RCELL TEMP=52.1 C Reaction cell temperature.
O2CELLTEMP 10 O2 CELL TEMP=50.2 C O2 sensor cell temperature.
BOXTEMP BOX TEMP=35.5 C Internal chassis temperature.
PMTTEMP PMT TEMP=7.0 C PMT temperature.
IZSDUTY IZS ON=0.00 SEC IZS temperature control duty cycle.
IZSTEMP IZS TEMP=52.2 C IZS temperature.
SO2 SO2= 261.4 PPB SO2 concentration for current range.
CO2 11 CO2= 0.00 PCT CO2 concentration.
O2 10 O2=0.00 WT% O2 concentration.
TESTCHAN T EST=3721.1 MV Value output to TEST_OUTPUT analog
output, selected with TEST_CHAN_ID
variable.
CLOCKTIME T IME=10:38:27 Current instrument time of day clock.
1 The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.
2 Engi neering software.
3 Current instrument units.
4 M100E.
5 M100E H.
6 M100E U.
10 O
2 option.
11 CO 2 option.