M100E MODBUS Register Map
MODBUS
Register Address
(dec., 0-based)
Description Units
MODBUS Floating Point Input Registers
(32-bit IEEE 754 format; read in high-word, low-word order; read-only)
0 PMT detector reading mV
2 1 Pre-amplified UV lamp intensity reading mV
4 UV lamp intensity reading mV
6 UV lamp ratio of calibrated intensity %
8 PMT electrical offset mV
10 UV lamp electrical offset mV
12 SO 2 slope for range #1
14 SO 2 slope for range #2
16 SO 2 offset for range #1 mV
18 SO 2 offset for range #2 mV
20 SO2 concentration for range #1 during zero/span calibration, just
before computing new slope and offset
PPB,
PPM 2
22 SO2 concentration for range #2 during zero/span calibration, just
before computing new slope and offset
PPB
24 SO 2 concentration for range #1 PPB
26 SO 2 concentration for range #2 PPB
28 12 SO 2 concentration for range #1, with O2 correction PPB
30 12 SO 2 concentration for range #2, with O2 correction PPB
32 Conc entration stability PPB
34 Stray light reading PPB
36 Reaction cell temperature C
38 1 IZ S temperature C
40 PMT temperature C
42 1, 2 Sample flow cc/m
44 Sampl e pressure “Hg
46 2 Vacuum pressure “Hg
48 Internal box temperature C
50 High voltage power supply output Volts
52 Diagnostic test input (TEST_INPUT_8) mV
54 Diagnostic temperature input (TEMP_INPUT_5) C
56 Diagnostic temperature input (TEMP_INPUT_6) C
58 Ground reference (REF_GND) mV
60 4096 mV reference (REF_4096_MV) mV
100 10 O
2 concentration %