Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation
A-32 05928 Rev F.0B
PRINTED DOCUMENTS ARE UNCONTROLLED DCN 5063
APPENDIX A-5: M100EU iDAS Functions, Revision F.0B
Table A-5: M100EU DAS Trigger Events, Revision F.0B
Name Description
ATIMER Automatic timer expired
EXITZR Exit zero calibration mode
EXITLS 2, 3 Exit low span calibration mode
EXITHS Exit high span calibration mode
EXITMP Exit multi-point calibration mode
EXITO2 10 Exit O2 calibration mode
SLPCHG Slope and offset recalculated
CO2SLC 11 CO2 slope and offset recalculated
O2SLPC 10 O2 slope and offset recalculated
EXITDG Exit diagnostic mode
PMTDTW PMT detector warning
UVLMPW UV lamp warning
DRKCLW 1 Dark calibration warning
RCTMPW Reaction cell temperature warning
O2TMPW 10 O2 sensor cell temperature warning
IZTMPW 1 IZS temperature warning
PTEMPW PMT temperature warning
SFLOWW Sample flow warning
SPRESW Sample pressure warning
VPRESW 2 Vacuum pressure warning
BTEMPW Box temperature warning
HVPSW High voltage power supply warning
1 M100E.
2 M100E H.
3 Low span option.
10 O
2 option.
11 CO 2 option.
Table A-6: M100EU iDAS Functions, Revision F.0B
Name Description Units
PMTDET PMT detector reading mV
PHABS 1 Pre-amplified UV lamp intensity reading mV
UVDET UV lamp intensity reading mV
LAMPR UV lamp ratio of calibrated intensity %
DRKPMT PMT electrical offset mV
DARKUV UV lamp electrical offset mV
SLOPE1 SO2 slope for range #1