Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation
05928 Rev F.0B A-31
DCN 5063 PRINTED DOCUMENTS ARE UNCONTROLLED
100 10 In O2 calibration mode
101 10 O
2 cell temperature warning
200 11 In CO2 calibration mode
MODBUS Coil Registers
(single-bit; read/write)
0 Maps to relay output signal 36 (MB_RELAY_36 in signal I/O list)
1 Maps to relay output signal 37 (MB_RELAY_37 in signal I/O list)
2 Maps to relay output signal 38 (MB_RELAY_38 in signal I/O list)
3 Maps to relay output signal 39 (MB_RELAY_39 in signal I/O list)
20 13 Triggers zero calibration of range #1 (on enters cal.; off exits cal.)
21 13 Triggers span calibration of range #1 (on enters cal.; off exits cal.)
22 13 Triggers zero calibration of range #2 (on enters cal.; off exits cal.)
23 13 Triggers span calibration of range #2 (on enters cal.; off exits cal.)
1 M100E.
2 M100E H.
10 O
2 option.
11 CO 2 option.
12 SO 2 with O2 correction option.
13 Set DYN_ZERO or DYN_SPAN variables to ON to enable calculating new slope or offset. Otherwise a calibration check
is performed.