Model 100EU Instruction Manual Addendum Appendix C – Warranty Repair Questionnaire
05930 Rev A C-1
Appendix C – Warranty Repair Questionnaire
CUSTOMER: _________________________________ PHONE: __________________________________________________
CONTACT NAME: ____________________________ FAX NO. __________________________________________________
SITE ADDRESS: _________________________________________________________________________________________
MODEL 100EU SERIAL NO.: ___________________ FIRMWARE REVISION: _____________________________________
1. ARE THERE ANY FAILURE MESSAGES? _______________________________________________________________
________________________________________________________________________________________________________
________________________________________________________________________________________________________
2. PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST
PARAMETERS BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT)
Parameter
*IF OPTION IS INSTALLED Recorded Value Acceptable Value
RANGE PPB/PPM
STABIL PPB 0.3 PPB WITH ZERO AIR
STABIL2 PPB <= 1 PPB WITH ZERO AIR
PRESS IN-HG-A AMBIENT ( - 2) IN-HG-A
SAMPLE FLOW cm3/MIN 650 ± 10%
PMT SIGNAL WITH
ZERO AIR mV -20 TO 150 mV
PMT SIGNAL AT
SPAN GAS CONC mV
PPB/PPM
0-5000 mV
0-20000 PPB
NORM PMT AT SPAN
GAS CONC mV
PPB/PPM
0-5000 mV
0-20000 PPB
UV LAMP mV 2000 TO 4000 mV
UV STAB mV < 15mV
LAMP RATIO mV 30 TO 120%
STR. LGT PPB 100 PPB/ ZERO AIR
DARK PMT mV 200-325
DARK LAMP mV -50 TO 200 mV
SLOPE 1.0 ± 0.5
OFFSET mV < 250 mV
HVPS V 400 – 800
RCELL TEMP ºC 50ºC ± 1
BOX TEMP ºC AMBIENT ± 5
PMT TEMP ºC 9.5ºC ± 2º CONSTANT
IZS TEMP* ºC 50ºC ± 1
ETEST mV 2000 mV ± 500
OTEST mV 2000 mV ± 1000
Values are in the Signal I/O
REF_4096_MV mV 4096mv±2mv and Must be Stable
REF_GND mV 0± 0.5 and Must be Stable