Operating Instructions Teledyne API - Model 200EH/EM Operation Manual
64
6.7.1. IDAS STRUCTURE
The iDAS is designed around the feature of a “record”, an automatically stored single data point. (e.g.
concentration, PMT signal level, etc.). Records are organized into data channels which are defined by
properties that characterize the:
Type of date recorded (e.g. concentration, PMT signal level, etc.);
Trigger event that causes the record to be made (e.g. every minute, upon exiting calibration mode,
etc.);
How many records to be stored, as well as;
How the information is to be stored (e.g. average over 1 hour, individual points, minimum value over
last 5 minutes, etc.).
The configuration of each iDAS channel is stored in the analyzer’s memory as a script, which can be edited from
the front panel or downloaded, edited and uploaded to the instrument in form of a string of plain-text lines
through the communication ports.
6.7.1.1. iDAS Channels
The key to the flexibility of the iDAS is its ability to store a large number of combinations of triggering events and
data parameters in the form of data channels. Users may create up to 20 data channels. For each channel one
triggering event is selected and one or all of the M200EH/EM’s 25 data parameters are allowed. The number of
parameters and channels is limited by available memory.
The properties that define the structure of an iDAS data channel are:
Table 6-7: iDAS Data Channel Properties
PROPERTY DESCRIPTION DEFAULT SETTING RANGE
NAME The name of the data channel. “NONE” Up to 6 letters or digits1.
TRIGGERING
EVENT
The event that triggers the data channel to
measure and store the datum ATIMER Any available event
(see Appendix A-5).
NUMBER AND
LIST OF
PARAMETERS
A User-configurable list of data types to be
recorded in any given channel. 1 - PMTDET Any available parameter
(see Appendix A-5).
REPORT PERIOD The amount of time between each channel data
point. 000:01:00
000:00:01 to
366:23:59
(Days:Hours:Minutes)
NUMBER OF
RECORDS
The number of reports that will be stored in the
data file. Once the limit is exceeded, the oldest
data is over-written.
100 1 to 1 million, limited by
available storage space.
RS-232 REPORT Enables the analyzer to automatically report
channel values to the RS-232 ports. OFF OFF or ON
CHANNEL
ENABLED
Enables or disables the channel. Allows a channel
to be temporarily turned off without deleting it. ON OFF or ON
CAL HOLD OFF Disables sampling of data parameters while
instrument is in calibration mode2. OFF OFF or ON
1 More with APICOM, but only the first six are displayed on the front panel).
2 When enabled records are not recorded until the DAS HOLD OFF period is passed after calibration mode. DAS HOLD OFF set in
the VARS menu (see Section 6.12.)
04521C (DCN5731)