Teledyne 311-D OPERATING TEMPERATURE RANGE 30 TO 125F, Ranges Of Analysis, Ppm Oxygen, 0-1000

Models: 311-D

1 18
Download 18 pages 49.01 Kb
Page 12
Image 12

OPERATING TEMPERATURE RANGE: 30° TO 125°F.

RESPONSE AND RECOVERY: AT THE SPECIFIED FLOWRATE (0.25 SCFH), THE SENSOR OUTPUT SHOULD BE, IN TEN SECONDS, WITHIN ±10% OF AN INTRODUCED GAS HAVING A CONCENTRATION OF 100 PPM OR MORE; FOR A CONCENTRATION OF LESS THAN 100 PPM, THE SENSOR OUTPUT SHOULD BE WITHIN ±10% IN SIXTY SECONDS (PROVIDED THAT THE SENSOR IS NOT ALREADY SUPERSATURATED WITH A VERY HIGHLY-CONCENTRATED GAS, AND THE INTRODUCED GAS IS NOT VERY LOW IN CONCENTRATION.

RANGES OF ANALYSIS: 0-100

PPM OXYGEN

0-1000

PPM OXYGEN

0-1

% OXYGEN

0-100

% OXYGEN (CAL.)

RECOMMENDED SPAN GAS: ATMOSPHERIC AIR.