319R

Appendix

 

 

 

 

 

 

 

 

 

Appendix

Specifications

 

 

Ranges:

0–200 PPM, 0–1999 PPM O2, and 0–

 

 

25% calibration range

 

Sensitivity:

1% of full scale

 

Signal Output:

Standard: 0–10 VDC, negative ground

 

 

4-20 mA, negative ground

 

 

plus contact closures for range

 

 

identification.

 

Display:

Digital 7 segment, 3½-digit LCD with

 

 

over range indicator.

 

Alarms:

One high alarm, adjustable & fail-safe.

 

 

One low alarm, adjustable & fail-safe.

 

 

One low alarm, fixed at 2% full scale and

 

 

fail-safe.

 

Accuracy:

±2% of full scale at constant temperature

 

 

and pressure of calibration

 

 

±5% of full scale throughout operating

 

 

temperature range (once temperature

 

 

equilibrium has been reached).

 

Response:

90% of full scale in 35 seconds.

 

Operating Temp. Range:

32 °F to 122 °F (0°C to 50°C)

 

Power:

85-264 VAC, 47-440 Hz or 10-36 VDC,

 

 

user specified, 10 W max.

 

Enclosure:

General purpose, panel mount

 

Cell Type:

Z-2

 

Cell Warranty:

6 months

 

Cell Life (Expected):

6 months

 

Dimensions:

2.75 "H × 6.0 "W × 2.75 "D

 

 

(63.5 mm × 152.4 mm × 63.5 mm)

 

 

Faceplate: 3.75 "H × 7.0 "W

 

 

(95.25 mm × 177.8 mm)

 

 

Probe unit: 4.0 "H × 6.0 "W × 3 "D

 

 

(101.6 mm × 152.4 mm × 76.2 mm)

 

 

 

TELEDYNE BROWN ENGINEERING

13

Analytical Instruments

Page 17
Image 17
Teledyne instruction manual 319R Appendix, Specifications