4.2.MAKING THE PNEUMATIC CONNECTIONS

Model GFC7000T Carbon Dioxide Analyzer

Table 4-1. Model GFC 7000T Pneumatic Connections

Rear Panel Label

Function

 

Connect a gas line from the source of sample gas here.

SAMPLE

Calibration gasses are also inlet here on units without

 

zero/span valve option installed.

 

 

EXHAUST

Connect an exhaust gas line of not more than 10 meters

long here.

 

 

 

PRESSURE SPAN

On units with zero/span valve option installed, connect a

 

gas line to the source of calibrated span gas here.

VENT SPAN

Not used

 

 

ZERO

On units with zero/span valve option installed, attach a

gas line to the source of zero air here.

 

 

 

 

This inlet supplies purge air to the GFC wheel housing

TO PURGE

Connect a source of dried air that has been scrubbed of

 

CO2.

 

This exhausts purge air to the GFC wheel housing.

FROM PURGE

Connect an exhaust gas line of not more than 10 meters

long here. It is only used on the GFC 7000TU instrument.

 

 

 

4.2.MAKING THE PNEUMATIC CONNECTIONS

CAUTION

Venting should be outside the shelter or immediate area surrounding the instrument.

1.Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2 meters long.

2.Attach sources of zero air and span gas

3.Span Gas is a gas specifically mixed to match the chemical composition of the type of gas being measured at near full scale of the desired measurement range.

When CO2 measurements are to be made with the Teledyne Instruments Model GFC 7000T Analyzer it is recommended that you use a gas calibrated to have a CO2 content equaling 80% of the range of compositions being measured.

Teledyne Analytical Instruments

5

Page 11
Image 11
Teledyne gfc 7000t manual Making The Pneumatic Connections, Model GFC7000T Carbon Dioxide Analyzer