Operating Instructions Teledyne API – T101 Operation Manual
98
4.6.5. ELECTRICAL TEST
The electrical test function creates a current, which substitutes the PMT signal, and feeds
it into the preamplifier board. This signal is generated by circuitry on the pre-amplifier
board itself and tests the filtering and amplification functions of that assembly along with
the A/D converter on the motherboard. It does not test the PMT itself. The electrical test
should produce a PMT signal of about 2000 ±1000 mV.
To activate the electrical test press:
SETUP X. X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SET UP
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X. X PRIMARY SETUP MEN U
CFG DAS RNGE PA SS CLK MORE EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
DIAG ELEC RANGE = 500.000 PPB O2=X.XXX
<TST TST> EXIT
Press NEXT until…
DIAG ELEC T RI CA L TEST
PREV NE XT ENTR EXIT
Press TST until…
DIAG ELE C PMT = 1732 MV SO2=X. XXX
<TST TST> EXIT
While the elec trical t est is
activa ted, PMT should equal:
2000 mV ± 10 00 mV
07266B DCN6485