4.6.6. LAMP CALIBRATION
An important factor in accurately determining H2S concentration, once the H2S is
converted to SO2, is the amount of UV light available to transform the SO2 into SO2* (see
Sections 10.1.1 and 10.1.2). The Model T101 compensates for variations in the intensity
of the available UV light by adjusting the H2S concentration calculation using a ratio
(LAMP RATIO)that results from dividing the current UV lamp (UV LAMP) intensity
by a value stored in the CPU’s memory (LAMP_CAL). Both LAMP Ratio and UV
Lamp are test functions viewable from the instrument’s front panel.
To cause the analyzer to measure and record a value for LAMP_CAL, press:
Exit at
any tim e
to return
to main
the
SETUP
menu
DIAG SIGNAL I / O
NEXT ENTR EXIT
Repeat Pressing NEXT until . . .
DIAG LAMP CALIBRATION
PREV NEXT ENT R EXIT
DIAG FCAL LAMP CAL VALUE:4262.4 m V
4 2 6 2 .4 ENTR EXIT
ENTR accepts the
new value
EXIT ignores the new
value
The value displayed is the
curre nt output of th e UV
source reference detector
SETUP X.X SECONDARY S ETUP MENU
COMM VARS DIAG EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
SAMPLE ENTER SET U P PASS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
07266B DCN6485