Getting Started Teledyne API – T101 Operation Manual
42
PMT
EXHAUS T T HROUGH OUTER
LAYER OF KICK ER
REAC T ION CELL PURGE
FLOW
SENSOR FLOW / PRESSURE
SENSOR PCA SAMPLE
FILTER
INSTRUMENT C HASSIS
EXH AUST GAS
OUTLET
VACUUM MANI FOLD
SAMPLE GAS
INLET
KICKER EX HA UST TO PUMP
PUMP
FLO W
CONTR OL
ASSY
SAMPLE
PRESSURE
SENSOR
UV
LAMP
HYDROCARB ON
SCRUBBER
(KICKE R)
SPAN GAS INLET
ZERO AIR I NLET
SAMPLE
CHAMBER
COM
NC NO
SO2
Scrubber
MOLYBDENUM
CONVERTER
SO2 H2S
H2S / SO2
MODE VALVE
Gas Flow when m ultigas versi on of
Analyzer is measuring SO
2.
Figure 3-13. Pneumatic Diagram of the T101 Standard Configuration
Table 3-8. H2S – SO2 Switching Valve Operating States
GAS
MODE CONDITION OF H2S –SO2 SWITCHING
VALVE VALVE PORT
CONNECTION
H2S Open to SO2 Scrubber and Molybdenum
Converter COM NO
SO2 Open to directly to Sample Chamber. Bypasses
SO2 Scrubber and Molybdenum Converter COM NC
H2S –SO2 Switches between above two states every 10
minutes. - -
1. Attach the 1/4" exhaust line to the EXHAUST port of the analyzer and to
the inlet port of the pump.