Teledyne API – T101 Operation Manual Getting Started
53
3.6. INITIAL CALIBRATION
3.6.1. BASIC CALIBRATION PROCEDURE
The following three-step procedure assumes that the instrument does not have any of the
available zero/span (Z/S) or IZS valve options installed. Section 6 contains instructions
for calibrating instruments with valve options.
The initial calibration should be carried out with the analyzer’s reporting range for
SINGLE range mode with a range span of 500 PPB (factory default settings for most
units). This will enable you to compare your results to the factory calibration.
STEP ONE: Set/verify the analog output reporting range of the T101:
SETUP X.X RANGE CONTROL MENU
MODE
SET UN IT EXIT
Press th is butt on to se t
the analyzer for
SNGL
DUAL or AU TO ranges
SETUP X.X RANGE: 500.0 CONC
0 0 5 0 0 .0 ENTR EXIT
To change th e value of the repo rting
range s pan, en ter the nu mber by
pressing the button under each digit
until t he expecte d value ap pears.
Press this button to sel ect th e
concentration units of measure:
PPB, PPM, UGM, MGM
EXIT ignores the new setting and
returns to the RANGE CONTROL
MENU.
ENTR accepts the new s etting and
returns to the
RANGE CONTROL MENU.
SETUP X.X
CFG DAS RNGE PASS CLK MORE EXI T
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
SETUP X.X RANG E: 500.0 Co n
c
0 0 0 5 0 .0 ENT R EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
07266B DCN6485