Getting Started Teledyne API – T101 Operation Manual
54
STEP TWO: Set the expected H2S span gas concentration.
This sequence causes the
analyzer to pro mpt for the
expect ed H2S s pan
concent rat ion.
M-P CAL H2S SPAN CONC: 450.0 Con
c
0 0 0 4 5 0 .0 ENTR EXIT
M-P CAL RANGE = 500.000 PPB H2S =X.XXX
< TST TST > ZERO CONC EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
The H2S span
concent ration v alues
automatically default to
450.0 Con c .
To change this value to
the actu al conc entrati on of
the spa n gas, en ter the
number by pressing the
button un der e ach digit
until th e expect ed value
appears.
EXIT ignores th e new settin g
and ret urns to the previ ous
display.
ENTR acc epts the n ew setti ng
and returns to the
previous display..
07266B DCN6485