Operating Instructions Teledyne API – T101 Operation Manual
60
TEMP Converter
Temperature
H2S present in the sample gas into SO2.
TEST2 Test Signal2 mV Signal of a user-defined test function on output channel A4.
TIME Clock Time hh:mm:ss The current day time for DAS records and calibration events.
1Shown as they appear when analyzer is in H2S mode. In SO2 mode appear as SO2 STB, SO2 OFFS & SO2 SLOPE. In
multigas mode, both versions appear.
To view the TEST Functions press the following touchscreen control button sequence:
1 Only appears if IZS option is
installed.
2 Only appears if analog output A4
is actively reporting a test function.
3 Shown as they appear when analyzer
is in H2S mode. In SO2 mode appear as SO2 STB, SO2 OFFS &
SO2 SLOPE . In multigas mode, both versions appear.
RANGE
H2S S T B3
SAMP FL
PRES
PMT
NORM PMT
UV LAMP
LAMP RATIO
STR. LGT
DARK PMT
DARK LAMP
H2S SLOPE3
H2S OFFS3
HVPS
RCELL TEMP
BOX TEMP
PMT TEMP
CONV TEMP
IZS TEMP1
TEST2
TIME
SAMPLE RANGE = 500.0 PPB SO2 400 PPB
< TST TST > CAL SET UP
Toggle <TST TST> buttons
to sc roll thr ou
g
h list of
Figure 4-1. Viewing T101 TEST Functions
NOTE
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range reading or the
analyzer’s inability to calculate it.
All pressure measurements are represented in terms of absolute pressure. Absolute, atmospheric
pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300 m gain in altitude. A variety of
factors such as air conditioning and passing storms can cause changes in the absolute atmospheric
pressure.