Operating Instructions Teledyne API – T101 Operation Manual
82
To access the DIAG functions press the following buttons:
SAMPLE ENTER SETUP PA SS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SET UP MENU
COMM VARS DIAG EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SE TUP
Within the COMM,
VARS and D IAG
menus, EXIT re turns
to the
SECONDARY
SETUP MENU
EXIT returns
to the PRIMARY
SETUP MENU
DIAG PRESSURE CALIBRATION
PREV NEXT ENTR EXIT
DIAG OPTIC TEST
PREV NEXT ENTR EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
DIAG AN ALOG OU T PUT
PREV NEXT ENTR EXIT
DIAG
A
NALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG TEST CHAN OUTPUT
PREV ENTR EXIT
DIAG ELECTRICAL TEST
PREV NEXT ENTR EXIT
DIAG LAMP CALIBRATION
PREV NEXT ENTR EXIT
EXIT returns
to th e mai n
SAMPLE
display
DIAG FLOW CALIBRATION
PREV NEXT ENTR EXIT
07266B DCN6485