Operating Instructions Teledyne API – T101 Operation Manual
84
4.6.2. ANALOG OUTPUT STEP TEST
This test can be used to check the accuracy and proper operation of the analog outputs.
The test forces all four analog output channels to produce signals ranging from 0% to
100% of the full scale range in 20% increments. This test is useful to verify the operation
of the data logging/recording devices attached to the analyzer.
To begin the Anog Output Step Test press:
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SE TUP
SAMPLE ENTER SETUP PA S S : 818
8 1 8 ENTR EXIT
SETUP X .X PRIM ARY SETUP MEN U
CFG DAS RNGE PASS CLK MORE EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
Exit-Exit
returns to the
DIAG menu
SETUP X .X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
DIAG
A
NALOG OUTPUT
PREV NEXT ENTR EXIT
DIAG AOUT ANALOG OUTPUT
0% EXIT
Perfor ms
analog output
step test.
0% - 100%
DIAG AOUT ANALO G OUTPUT
[0%] EXIT
Pressi ng the but ton unde r “0%” whil e perfor ming the t est will pa use
the tes t at that l evel. Bracket s will appear aroun d the value; example :
[20%] P ressin g the same bu tton ag ain will resume t he test.
07266B DCN6485