2. SPECIFICATIONS AND APPROVALS
2.1. T200H/M OPERATING SPECIFICATIONS
Table 2-1: Model T200H/M Basic Unit Specifications
Min/Max Range
(Physical Analog Output) T200H: Min: 0-5 ppm; Max: 0-5000 ppm T200M: Min: 0-1 ppm; Max: 0-200 ppm
Measurement Units ppm, mg/m3 (user selectable)
Zero Noise <20 ppb (RMS)
Span Noise <0.2% of reading above 20 ppm
Lower Detectable Limit 40 ppb (2x noise as per USEPA)
Zero Drift (24 hours) <20 ppb (at constant temperature and voltage.)
Zero Drift (7 days) <20 ppb (at constant temperature and voltage.)
Span Drift (7 Days) <1% of reading (at constant temperature and voltage.)
Linearity 1% of full scale
Precision 0.5% of reading
Lag Time 20 s
Rise/Fall Time 95% in <60 s (~10 s in NO only or NOX only modes)
T200H:
40 cm³/min sample gas through NO2
converter & sensor module
250 cm3/min ± 10% through bypass
manifold
290 cm³/min total flow
T200M:
250 cm³/min sample gas through NO2
converter & sensor module
Gas Flow Rates
O2 Sensor option adds 80 cm³/min to total flow though T200H/M when installed.
Temperature Range 5 - 40 C operating range
Humidity Range 0-95% RH non-condensing
Dimensions H x W x D 18 cm x 43 cm x 61 cm (7" x 17" x 23.6")
Weight, Analyzer 18 kg (40 lbs) without options
Weight, Ext Pump Pack 7 kg (16 lbs)
AC Power
T200H:
100V-120V, 60 Hz (175W)
220V-240V, 50 Hz (155W)
T200M:
100V-120V, 60 Hz (55W)
220V-240V, 50 Hz (75W)
Power, Ext Pump 100 V, 50 Hz (300 W); 100 V, 60 Hz (255 W); 115 V, 60 Hz (285 W);
220 - 240 V, 50 Hz (270 W); 230 V, 60 Hz (270 W)
Environmental Installation category (over-voltage category) II; Pollution degree 2
Analog Outputs 4 user configurable outputs
Analog Output Ranges
All Outputs: 0.1 V, 1 V, 5 V or 10 V
Three outputs convertible to 4-20 mA isolated current loop.
All Ranges with 5% under/over-range
Analog Output Resolution 1 part in 4096 of selected full-scale voltage (12 bit)
Status Outputs 8 Status outputs from opto-isolators, 7 defined, 1 spare
Control Inputs 6 Control inputs, 4 defined, 2 spare
Alarm outputs 2 relay alarms outputs (Optional equipment) with user settable alarm limits
- 1 Form C: SPDT; 3 Amp @ 125 VAC
Standard I/O
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital outputs
6 opto-isolated digital inputs
4 analog outputs
Optional I/O
1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs
07270B DCN6512