Teledyne API - Model T200H/T200M Operation Manual Getting Started
57
3.5.3.3. Internal Flow for O2 Sensor Option 65A
Please see Section 3.7.2 for calibration connections and method.
VACUUM
PRESSURE
SENSOR
SAMPLE
PRESSURE
SENSOR
O3 FLOW
SENSOR
FLOW PRESSURE
SENSOR PCA
INSTRUMENT CHASSIS
NO/NOX
VALVE
AUTOZERO
VALVE
O3
Destruct
O3
GENERATOR
EXHAUST
GAS
OUTLET
SAMPLE
GAS
INLET
EXHAUST MANIFOLD
O2
Sensor
Orifice Dia.
0.007"
Orifice Dia.
0.004"
PUMP
NOXExhaust
Scrubber
Orifice Dia.
0.007"
REACTION
CELL
PMT
PERMAPURE
DRYER
Filter
Orifice Dia.
0.004"
OPTION, O2
SENSOR, P/N 04453
NC
COMNO
NC
COM
NO
NO2
Converter
Figure 3-23: T200H – Internal Pneumatics with O2 Sensor Option 65A